MAGNETIC RESONANCE IMAGING MEETS NEONATOLOGY Download the brochures and FAQ document
LMT is the first company offering
an MR-compatible incubator
with integrated coil for newborns.
It facilitates safe and convenient
transport from the NICU to the MRI
department with an MR-compatible
trolley, MR-compatible gas and
power supply and necessary
compatible accessories.
The MR Diagnostics Incubator System nomag® IC offers the solution for a large, thus far unsolved
problem: with the incubator system, pre and term born babies can be examined for the first time after
birth by optimal non-invasive diagnostics with magnetic resonance imaging...
A wide range of accessories enhances the optimum and efficient operation of this system, such as: MR
compatible ventilation, integrated monitoring ( SpO2 ), graphic display,body coil or trolley for ambulance
transport.
The system is compatible with the following MRI scanners: Scanner with 1.5 Tesla field intensity
Siemens: Sonata, Symphony, Avanto, Espree - Philips: Intera, Achieva - GE: Signa
Scanner with 3.0 Tesla field intensity
Siemens: Trio, TIM Trio, Verio. - Philips: Intera, Achieva. - GE: Signa.
![]() CHUV, Lausanne.
The LMT nomag® IC 1.5 / 3.0 MR-compatible incubator provides a stable climate with bactericidal
humidification as required by premature infants and newborns weighing up to approx. 4.500g (10 lbs)
and a max. body length of approx. 55 cm. Thanks to its construction, it supports safe handling and
operation during the MRI and intra-hospital transport.
Technical Details
Patient
Weight : max. 4.5 kg
Size : max. 55 cm
Head circumference : max. 38 cm
Measurements
Entire device : 2100 x 760 x 1000 mm³ (LxWxH)
Incubator : 1870 x 350x 350 mm³ (LxWxH)
Weight
Incubator : max. 475 W
Power consumption UPS : max. 700 W
Climate
Humidification : 30-70 %rH closed loop control
Air speed : < 10 cm/s
Fresh air supply : 15 l/min
Air temperature regulation : 28 °C - 39 °C (82.4°F - 102.2°F)
Warm-up time : 20 min from 20°C to 32°C (69.8°F - 87.8°F)
Download the FDA clearance offical document
Please find an overview of the publications here: Elspeth H. Whitby et al.
“Ultrafast Magnetic Resonance Imaging of the Neonate in a Magnetic Resonance-Compatible M. Feijen, R N, M.J.G. Plasmans, R N Rangmar Goelz, Eva Bültmann, Thomas Nägele
"First Expereinces with Magnetic Resonance Imaging of Neonates Using a Megnetic-Resonance- Dr. Katrin Klebermass, Dr. Peter Brugger, Prof. Arnold Pollak, Prof. Daniela Prayer, MD Zsófia Rona, Prof. Manfred Weninger
"Comparison of neonatal MRI examinations with and without an MR-compatible incubator".
|